Analisador Bancada por UV visível OMA-406

Classic OMA performance in a lab-friendly form factor.
  • Características
  • Aplicações
  • Especificações
  • Download
  • Functionally and electronically identical to the OMA-300 UV-VIS, the solid state OMA-406R is designed to be rack-mounted. This turn-key unit fits a standard 19" rack, providing typical OMA outputs including 4-20mA, RS232, and Modbus (TCP/IP, Ethernet, serial).

    The OMA-406R is often more suitable for laboratory environments but is occasionally installed in shelters in the field. This system can be used with any AAI flow cell and/or sampling system.
  • Simply a rack-mounted version of the OMA-300 UV-VIS, the OMA-406R is designed for virtually all of the same applications. AAI engineers know that each process has highly specific needs, so we made our product line versatile and our instruments primed for integration.

  • General Performance

    Measurement Technology

    nova-II UV-VIS diode array spectrometer

    Light Source

    pulsed xenon lamp (~5 year lifespan); deuterium and tungsten sources available

    Sample Introduction

    flow-through cell; standard or custom-design sampling system (optional)

    Repeatability

    ±0.5 % of scale

    Photometric Accuracy

    ±0.004 AU

    Calibration

    for many applications, factory calibrated with certified calibration gases; for safety reasons, some applications require on-site calibration

    Verification

    easy verification/validation with standard certified gas samples and neutral density filters


    Operating Conditions

    Ambient Temperature

    standard: 0 to 55 °C (32 to 131 °F) 
    optional: -20 to 55 °C (-4 to 131 °F)

    Sample Conditions

    Temperature

    immersion probe: -20 to 150 °C (-4 to 302 °F) 
    flow-through cell: -20 to 150 °C (-4 to 302 °F)

    Pressure

    immersion probe: 100 bar (1470 psig) 
    flow-through cell: 206 bar (3000 psi)

    Environment

    indoor/outdoor (no shelter required)


    Hardware

    Size

    6.54" H x 16.63" W x 10.13" D (166mm H x 422mm W x 257mm D) 
    optional basic sampling system: 24" H x 30" W x 8" D (610mm H x 760mm W x 200mm D)

    Weight

    20 lbs. (9 kg)

    Wetted Materials

    analyzer: stainless steel 316/316L; K7 glass; Viton 
    optional basic sampling system: stainless steel 316/317; quartz; Viton

    Outputs

    one galvanically isolated 4-20mA output per component; modbus TCP/IP (optional); RS232 (optional); Fieldbus, Profibus, and HART (all optional); two digital outputs for fault and sampling system control (user programmable)

    Electrical Requirements

    85 to 264 VAC 47 to 63 Hz

    Power Consumption

    45 watts

    Area Classification

    General Purpose (standard) / Class I, Div. 2 (optional) / Class I, Div. 1 (optional) / ATEX Exp II 2(2) GD (optional)

     